7 Testing and Maintenance
90
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
Evaluation
System
(Sys 1-2)
Perform the system
test according to sub
clause 19.201 of
IEC 60601-
1:2005+A1:2012/
EN 60601-
1:2006+AC:2010, if
applicable, after
forming a system.
Equipment Leakage Current:
Sys1 <= 100 µA (Normal
Condition)
Sys2 <= 300 µA (Single Fault
Condition)
Sys:PSys1/PSys2 or
Sys:FSys1/FSys2
System
(Sys 3)
Perform the system
test according to sub
clause 19.201 of
IEC 60601-
1:2005+A1:2012/
EN 60601-
1:2006+AC:2010, if
applicable, after
forming a system.
Protective Earth Leakage
Current if medical electrical
system components are
connected to the same Multiple
Portable Socket-Outlet:
Sys3 <= 300 µA
Sys: PSys3 or
Sys: FSys3
Key: P = Pass, F = Fail, X = test result value to be recorded
Test Block
Test or Inspection
to be Performed
Expected Test Results
Record the Results (mandatory for
Philips Personnel only)
What to Record
Actual Results
Yes
No
Safety and Functional Test passed
Repair required at a later date, safety and functional test passed
Device must be taken out of operation until repair and passed tests
Device failed and must be taken out of operation.
Notes:
Next Recurrent Test:
Name:
____________________________________________________
Date/Signature:
_____________________________________________
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...