11 Upgrades
220
9
Fit the new connector cover that comes with the upgrade kit as described in step 8 for Option B71
on
10
Replace the reassembled front end assembly as described in “Setting the Fetal Recorder Offset” on
page 97.
11
Fit the SpO
2
board holder to the bottom housing with two screws from the upgrade kit.
12
Next, fit the SpO
2
board as described in “Replacing the SpO2 Assembly” on page 196.
13
Then refit the noninvasive blood pressure assembly as described in “Replacing the Noninvasive
Blood Pressure Assembly” on page 192.
14
Replace the top cover assembly as described in “Replacing the Top Cover Assembly” on page 180.
15
For tests to perform after upgrading, see “When to Perform Tests” on page 56.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...