7 Testing and Maintenance
85
Expected Test Results
SpO2 Performance Test
This test checks the performance of the SpO
2
measurement.
Tools required: none
1
Connect an adult SpO
2
transducer to the SpO
2
connector.
2
Measure the SpO
2
value on your finger (this assumes that you are healthy).
3
The value should be between 95% and 100%.
Expected Test Results
Measurement Validation
The SpO
2
accuracy has been validated in human studies against arterial blood sample reference
measured with a CO-oximeter. In a controlled desaturation study, healthy adult volunteers with
saturation levels between 70% and 100% SpO
2
were studied.
The population characteristics for those studies were:
•
about 50% female and 50% male subjects
•
age range: 19 to 39
•
skin tone: from light to dark brown
NOTE
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.
However, it can be used to demonstrate that a particular pulse oximeter monitor reproduces a
calibration curve that has been independently demonstrated to fulfill a particular accuracy
specification.
Pulse rate accuracy has been validated with an electronic pulse simulator.
Test
Expected test results
Accuracy test
x1 = value displayed by monitor
Difference
≤
3 mmHg
Leakage test
x2 = leakage test value
x2 <6 mmHg
Linearity test
x3 = value displayed by monitor
Difference
≤
3 mmHg
Valve Test
x4 = value <10 mmHg
Test
Expected test results
SpO
2
Performance Test
95%—100%
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...