3 Installation Instructions
32
5
Insert the probe covers into the base station. Up to two units of 16 probe covers each can be
stored within the base station.
6
Insert the thermometer into the base station until it clicks into place.
7
Insert the other end of the communication cable into the MIB/RS232 (1) port of the monitor.
8
Configure the MIB/R232 port at the fetal monitor to iTemp.
Switch to the Service Mode.
Select the
Main Setup
SmartKey to open the
Main Setup
window and select
Hardware
from the
menu.
In the
Setup Hardware
window select
Interfaces
.
In the
Setup Interfaces
window select
iTemp
for the MIB/R232 port.
NOTE
When disconnecting the thermometer cable from the base station, DO NOT twist the connector. Pull
on the ring of the connector to disconnect the cable.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...