12 Specifications
227
1
The power loss indication functionality of the Nurse Call Relay board is not supported with fetal
monitors.
Performance Specifications
NOTE
Your monitor's default settings can be permanently changed in Configuration Mode.
The default settings specified here refer to the settings initially shipped with the monitor.
Fetal Monitors
RS232 (Standard)
Connectors
RJ45 (8-pin)
Power none
Insulation
Basic insulation (reference voltage: 250 V; test voltage: 1500 V)
RS232 (Independent display
interface option)
Connectors
RJ45 (8-pin)
Power
none
Isolation
none
Flexible Nurse Call
Relay
1
Connectors
20 pin MDR (Mini D-Ribbon), active open and closed contacts
Contact
<=100 mA, <=24 V DC
Isolation
Basic insulation (reference voltage: 250 V; test voltage: 1500 V)
Delay
<[Configured L0.5] sec
Interface Specifications
Fetal Monitor Performance Specifications
Alarm Signal
System alarm delay.
The system alarm delay is the processing
time the system needs for any alarm to be
indicated on the fetal monitor, after the
measurement has triggered the alarm.
less than 4 seconds
Pause duration
1, 2, 3 minutes or infinite, depending on the
configuration
Extended alarm pause
5 to 10 minutes
Sound pressure range
min. 0 dB(A) max. 45—85 dB(A)
Review Alarms
Information
all alarms/INOPs, main alarms on/off, alarm
silence and time of occurrence
Capacity
300 items
Real time Clock
Range
from: January 1, 1997, 00:00 to:
December 31, 2080, 23:59
Accuracy
better than ±1 min. per month
Hold Time
infinite if powered by AC; otherwise at least
48 hours
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...