11 Upgrades
218
Option B72 (FM30 only)
1
Remove the top cover assembly as described “Removing the Top Cover Assembly” on page 141.
2
Remove the noninvasive blood pressure assembly.
3
Remove the front end assembly -- see “Removing the Front End Assembly” on page 200.
4
Remove the connector block frame from the front end assembly (see “Disassembling the Front
End Assembly” on page 202). Leave the fetal sensor socket connector block in place.
5
Remove the blood pressure connector from the bus master board.
6
Join the new SpO
2
connector to the noninvasive blood pressure connector.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...