8 Troubleshooting
115
Alarm Tones
Alarm Behavior
If your monitor did not alarm in the way in which the end user expected it to, please consult the
Instructions for Use for possible setup issues, or configuration settings which could affect alarm
behavior.
Fetal Recorder
LAN cable connected, IP address assigned, monitor connected to OB TraceVue/
IntelliSpace Perinatal.
OB TraceVue/IntelliSpace Perinatal may send a prompt message, indicating a possible
problem.
Icon
Explanation
Symptoms
Possible Causes of Failure
Failure Isolation and Remedy
INOP Message
Speaker
Malfunct
is displayed
Speaker cable disconnected
Reconnect the speaker cable.
Speaker defective
Replace the speaker.
Sound amplifier on the Main CPU Board
defective
Replace the Main CPU Board.
An alarm occurs, but no
alarm sound is issued.
Volume is set to 0
Increase the volume.
Speaker defective
Replace the speaker.
Sound amplifier on the Main CPU Board
defective
Replace the Main CPU Board.
Symptom
Possible Cause
Corrective Action
A Paper empty warning is
issued in the status line at
the bottom of the screen,
but paper is not out.
The drawer is open.
Close the drawer.
A paper jam occurred.
Open the drawer, remove paper, tear off the
crumpled paper and re-load, or load a new
pack of paper. Close the drawer.
The paper sensor is dirty.
Clean the paper sensor.
The paper sensor is defective.
Exchange the paper sensor -- see “Repair
and Disassembly” on page 139 and “Parts”
on page 125.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...