12 Specifications
236
Tympanic Temperature
Complies with:
•
EN 12470-5 (Clinical thermometers - Part 5:2003: Performance of infra-red thermometers)
•
ASTM E1965-98 (Infrared Thermometers for Intermittent Determination of Patient
Temperature)
with minor exceptions as noted below.
The fetal monitor additionally complies with ISO 80601-2-56:2009 / EN ISO 80601-2-56:2012
Bradycardia
Difference to low limit 0—50 bpm 5 bpm steps
max. 14 seconds
Clamping at 30—100 bpm
5 bpm steps
SpO
2
Alarm Specifications Range
Adjustment
Delay
Performance Specifications
Temperature Resolution
0.1°C or 0.1°F
Response Time
less than 2 seconds
Temperature Calibrated Accuracy Specifications (out of the Factory)
Ambient Temperature
Target Temperature
Accuracy
25.0°C (77.0°F)
37.7°C—38.9°C (98.4°F—102.0°F)
±0.1°C (±0.2°F)
16.0°C—33.0°C (60.8°F—
91.4°F)
33.0°C—42.0°C (91.4°F—107.6°F)
±0.2°C (±0.4°F)
Temperature Calibrated Accuracy Specifications (after recalibration using Genius 2 Checker/Calibrator)
Ambient Temperature
Target Temperature
Accuracy
16.0°C—33.0°C (60.8°F—
91.4°F)
36.0°C—39.0°C (96.8°F—102.2°F)
±0.2°C (±0.4°F)
16.0°C—33.0°C (60.8°F—
91.4°F)
<36.0°C or >39.0°C
(<96.8°F or >102.2°F)
±0.3°C (±0.5°F)
ASTM laboratory requirement for IR thermometers in the display range 37.0°C—39.0°C (98.0°F—102.0°F) is
±0.2°C (±0.4°F)
, whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM standards
E667-86 and E1112-86 is
±0.1°C (±0.2°F)
.
Clinical accuracy characteristics and procedures are available from Covidien llc on request. To verify the accuracy, use
a certified black body as specified in EN ISO 80601-2-56, Annex C, or use a Genius 2 Checker/Calibrator - available
from Covidien llc under part number 303097.
Clinical repeatability: meets section A.5 of EN ISO 80601-2-56(E) per Covidien llc technical report. Data is available
from Covidien llc on request.
Displayed Temperature Measurement Range
Mode
Range °C
Range °F
Ear
33.0—42.0
91.4—107.6
Oral (ear + 0.6°C)
33.6—42.0
92.5—107.6
Core (ear + 1.04°C)
34.0—42.0
93.2—107.6
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...