12 Specifications
224
WARNING
Explosion Hazard:
Do not use in the presence of flammable anesthetics, such as a flammable
anesthetic mixture with air, oxygen, nitrous oxide, or in oxygen rich environment. Use of the devices in
such an environment may present an explosion hazard.
Physical Specifications
Fetal Monitor
Fetal Monitor Sounds
Tympanic Temperature
Operating Temperature Range
16ºC—33ºC (60.8ºF—91.4ºF)
Monitor Physical Specifications
M2702A/M2703A
Power
Supply Voltages
100 VAC—240 VAC ±10%
Supply Frequency Range
50 Hz—60 Hz
Power Consumption
(current)
0.7—0.4 A (M2702A/M2703A)
1.3—0.7 A (M8023A#E25
Dimensions and
Weight
Size (without options) mm/
(in):
width x height x depth
286 x 133 x 335 mm ±1%
(11.3 x 5.2 x 13.2 in ±1%)
Weight
<5.1 kg (11.2 lbs)
Degree of Protection Against Electrical Shock
Type CF
Electrical Class
Class II equipment
Electrical Power Source
External (AC) without battery option #E25
Internal (LiIo battery) if with battery option #E25
Mode of Operation
Continuous operation
Water Ingress Protection Code
IP X1 (provided recorder drawer is shut)
Global Speed
6.25 mm/sec, 12.5 mm/sec, 25 mm/sec, 50 mm/sec
Startup Time
Time taken from switching
on the monitor to seeing the
first parameter labels
<30 seconds
Source
Description
Patient alarms and INOPs
See the sections on Patient Alarms and INOPs, Standard Philips Alarms, and
ISO/IEC Standard Audible Alarms in the Instructions for Use.
Ultrasound Doppler
Direct transmission of Doppler echoes to the speaker of the fetal monitor.
Pulse from SpO
2
, MECG, DECG
QRS tone
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...