10 Repair and Disassembly
140
Tools Required
CAUTION
When replacing the front cover, do not over-torque the screws. Excessive torque may damage the
plastic screw mountings.
You need the following tools:
Screws Used
The following picture shows the range of screws used in the Avalon FM20/30/40/50 fetal monitors:
Flat-head screwdriver, head thickness 0.5 mm to fit
transducer screw
Torx-head screwdriver, size T-10, minimum shaft length
80 mm
Small flat-head screwdriver, 2.0-3.0 mm
Long-nosed pliers
Screw Ejot K30x8, T-10
Screw Ejot 3x10, T-10
Screw Ejot 3x8, T-10
Screw Ejot 2.5x8, T-8
Screw M3x12 with washer, T-10
Screw M3x10 with washer, T-10
Screw M3x8 with washer, T-10
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...