12 Specifications
235
SpO
2
Performance Specifications
SpO
2
The specified accuracy is
the root-mean-square
(RMS) difference
between the measured
values and the reference
values
Range
0—100%
Accuracy
Philips Reusable Sensors:
M1191A/B, M1191AL/BL, M1191ANL, M1192A, M1192AN = 2%
(70%—100%)
M1191T, M1192T, M1194A, M1194AN, M1196A, M1196T = 3%
(70%—100%)
Philips Disposable Sensors with M1943A(L):
M1131A, M1901B, M1903B, M1904B = 3% (70%—100%)
M1133A, M1134A = ±2% (70%—100%)
NellcorPB® Sensors with M1943A(L):
MAX-A, MAX-AL, MAX-P, MAX-N, D-25, D-20, N-25, OxiCliq A,
P, N = 3% (70%—100%)
Masimo Reusable Sensors® with LNOP MP12 or LNC MP10:
LNOP DC-I, LNOP DC-IP, LNOP YI, LNCS DC-I, LNCS DC-IP:
2% (70%—100%)
LNOP TC-I, LNCS TC-I: 3.5% (70%—100%)
Masimo Disposable Sensors® with LNOP MP12 or LNC
MP10:
LNOP Adt, LNOP Adtx, LNOP Pdt, LNOP Pdtx, LNCS Adtx,
LNCS Pdtx: 2% (70%—100%)
LNOP Neo-L, LNCS Neo-L: 3% (70%—100%)
Resolution
1%
Pulse
Range
30—300 bpm
Accuracy
±2% or 1 bpm, whichever is greater
Resolution
1 bpm
Sensors
Wavelength range
500—1000 nm.
Information about the wavelength range can be especially useful to
clinicians (for instance, when photodynamic therapy is performed).
Emitted Light
Energy
≤
15mW
Pulse Oximeter Calibration Range
70%—100%
SpO
2
Alarm Specifications Range
Adjustment
Delay
SpO
2
50—100%
1% steps
(0, 1, 2, 3,... 30) +
4 seconds
Desat
50—Low alarm limit
1% steps
Pulse
30—300 bpm
1 bpm steps (30—40 bpm)
5 bpm steps (40—300 bpm)
max. 14 seconds
Tachycardia
Difference to high limit 0—50 bpm 5 bpm steps
max. 14 seconds
Clamping at 150—300 bpm
5 bpm steps
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...