7 Testing and Maintenance
89
Noninvasive
Blood Pressure
Performance
Tests
Perform the
Accuracy Test -- see
“Accuracy Test” on
page 83.
X1 = value displayed by monitor
Difference <= 3 mmHg
PN:P/X1 or
PN:F/X1
Performance
Leakage Test -- see
“Leakage Test” on
page 84.
X2 = leakage test value
X2 < 6 mmHg
PN:P/X2 or
PN:F/X2
Performance
Linearity Test -- see
“Leakage Test” on
page 84.
X3 = value displayed by monitor
Difference <= 3 mmHg
PN:P/X3 or
PN:F/X3
Performance Valve
Test -- page
“Leakage Test” on
page 84.
X4 = value < 10 mmHg
PN:P/X4 or
PN:F/X4
SpO
2
Performance
Test
Perform the SpO
2
Performance Test --
see “SpO2
Performance Test”
on page 85.
Value should be between 95%
and 100%
No reporting necessary
Safety (2)
Perform Safety Test
(2): Sum of
Functional Earth
and Equipment
Leakage Current -
Normal Condition.
NC with mains cable:
Maximum leakage current (X1)
≤
100 µA
S(1):P/X1 or
S(1):F/X1
Safety (3)
Perform Safety Test
(3): Sum of
Functional Earth
and Equipment
Leakage Current -
Single Fault
Condition (Open
Earth).
SFC with mains cable:
Maximum leakage current (X2)
≤
100 µA
S(2):P/X2 or
S(2):F/X2
Safety (4)
Perform Safety Test
(4): Applied Part
Leakage Current -
Single Fault
Condition, mains on
applied part.
Maximum leakage current (X1)
≤
50 µA
S(3):P/X1 or
S(3):F/X1
Test Block
Test or Inspection
to be Performed
Expected Test Results
Record the Results (mandatory for
Philips Personnel only)
What to Record
Actual Results
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......