5 Interfaces
50
NOTE
Other USB devices, e.g. USB sticks, iPods etc. are not supported by the USB IF board.
Do not
use
USB adapters to connect PS/2 or other devices to the USB board.
Flexible Nurse Call Relay
The Flexible Nurse Call board contains 2 connectors. A phone jack type connector and a multi-port
connector. The phone jack type connector has a single close-on-alarm relay. The multiport connector
has three alarm relays which are configurable to be open or closed on alarm.
Connections
Flexible Nurse Call Relay Connections at Monitor
See “Multi-Port Nurse Call Connector Test (Flexible Nurse Call)” on page 86 for details on the
Flexible Nurse Call relay connector pin assignment.
Nurse Call Relay
Connectors
Contact
Isolation
Flexible Nurse Call Relay 20 pin MDR (Mini D-Ribbon),
active open and closed contacts,
3.5 mm phone jack, active closed
contact only
≤
100 mA,
≤
24 VDC
1.5 kV
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...