8 Troubleshooting
118
LAN / RS232
Trace is not printed
correctly with a reference
to the paper gridlines.
The Offset needs adjusting.
Calibrate the recorder and change the offset
-- see “Setting the Fetal Recorder Offset”
on page 97.
The wrong paper scale is being used.
Ensure the paper you are using matches the
paper scale setting.
Symptom
Possible Cause
Corrective Action
Symptoms
Cause of Failure
Failure Isolation and Remedy
External device (such as a
surveillance system like
OB TraceVue/
IntelliSpace Perinatal) not
receiving data.
The LAN/RS232 port is not configured for
data export.
Check the configuration of the LAN/RS232
ports in Configuration Mode.
The cable between the external device and
the monitor is not connected correctly or
defective.
Check the cable and replace it if necessary.
The external device does not support the
version of the data export protocol used in
the monitor.
Check if the device supports the version of
the data export protocol. Upgrade the
device or the monitor if necessary (if
matching versions exist).
A terminal concentrator is used in between
the device and the monitor, and a protocol
with dynamic speed negotiation is used.
Some terminal concentrators do not
support changing the transmission speed
(baud rate) dynamically. Check if the
connection works without the concentrator.
The LAN/RS232 board is in a wrong slot
(slot has been changed after software
configuration or an additional board has
been plugged in).
Verify correct the placement of the I/O
boards.
The LAN/RS232 board is defective.
Check the board and replace it if necessary.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...