6 Connection to a Network
53
Equipment Label and OB TraceVue/IntelliSpace Perinatal Fetal
Monitor Domain Name
For connection to an OB TraceVue/IntelliSpace Perinatal system over a network, OB TraceVue/
IntelliSpace Perinatal requires each fetal monitor to have a unique equipment label.
When a fetal monitor is configured with an equipment label, this equipment label is sent as part of the
CI message. In OB TraceVue/IntelliSpace Perinatal, it is possible to specify a Fetal Monitor Domain
Name in the fetal monitor configuration user interface (see OB TraceVue/IntelliSpace Perinatal
Instructions for Use for details). The OB TraceVue/IntelliSpace Perinatal system compares this name
with the fetal monitor equipment label. Only if the domain name matches the beginning of the
equipment label string, is the CI message processed and accepted. Otherwise the CI message is
ignored.
Using this filtering process, you can avoid conflicts where there are multiple OB TraceVue/
IntelliSpace Perinatal systems operating in the same network segment by controlling which monitors
connect to a specific OB TraceVue/IntelliSpace Perinatal system, and which monitors are excluded.
To enter or change the equipment label:
1
Enter Configuration Mode.
2
Select
Main Setup.
3
Select
Bed Information
.
4
In the
Bed Information
menu, select
Equipment Label
.
5
Enter the desired equipment label for your monitor (up to 12 alphanumeric characters are
displayed), then press
Enter
.
Data Export/ Data Out Functionality
Data output/export is possible via LAN, or MIB/R232 interfaces.
The I/O boards typically provide data to externally attached devices, for example to RS232 based data
collection devices.
For a multi-purpose card it is necessary to configure the card for a particular purpose first. The MIB/
RS232 card can support Tympanic Temperature, data import, data export, external touch display
(FM40/FM50 only).
For more detailed information on this subject please refer to the Data Export Interface Programming
Guide.
Configuring the LAN Data Export Setting
The data that can be exported via the LAN interface is configurable. You can choose between the
following options: all, anonymous data, off. In case of anonymous data the patient name and given
name are not included in the data stream.
1
To change the
CentralMonitorng
configuration switch, first switch to Configuration Mode.
2
To configure the
LAN Data Export
setting, in Configuration Mode, select
Main Setup
to enter the
Main Setup
menu.
3
Select
Global Settings
.
4
Select
LAN Data Export
and toggle the appropriate setting.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...