7 Testing and Maintenance
57
Other Regular Tests and Tasks
Repairs
Upgrades
Tympanic Temperature performance testing
Tympanic Temperature calibration
Service Event
Test Block(s) Required - Complete these tests
Service Event
Test Block(s) Required - Complete these tests
Visual Inspection
Perform Visual Inspection
Transducer and Patient Module Testing
See
Recorder Maintenance
Regular cleaning and maintenance (see “Transducers and
Patient Modules; Functional Tests” on page 91)
Perform the recorder selftest (see “Fetal Recorder Selftest
Report” on page 98)
Service Event
Test Block(s) Required - Complete these tests
Repairs when the monitor has been damaged by impact,
liquid ingression, fire, short circuit, or electrical surge
Perform Visual Inspection, Power On, all Safety tests, and
Full Performance Assurance tests
Repairs where the power supply, the mains socket or an
interface board is removed, or replaced, or the protective
earth ground connection is disrupted
Perform Visual Inspection, Power On, all Safety Tests,
and Basic Performance Assurance tests
Repairs where the NBP pump has been replaced
Perform Visual Inspection, Power On Test, all Safety tests,
Basic Performance Assurance test, and NBP Performance
test and calibration
All repair events involving any of the following:
Noninvasive Blood Pressure connector, SpO
2
connector/
board or the red fetal sensor sockets
Perform Visual Inspection, Power On, Full Performance
tests, and all Safety tests
Service Event
Test Block(s) Required - Complete these tests
Software upgrades
Perform Visual Inspection, Power On test, and Basic
Performance Assurance test unless otherwise specified in
the Upgrade Installation Notes shipped with the upgrade
Hardware upgrades
Perform Visual Inspection, Power On test, and Basic
Performance Assurance test unless otherwise specified in
the Upgrade Installation Notes shipped with the upgrade
Installation of interfaces or hardware where the power
supply, or parameter boards need to be removed
Perform Visual Inspection, Power On test, Basic
Performance tests, and all Safety tests
Noninvasive Blood Pressure hardware upgrade M2702AU
Option B71
Perform Visual Inspection, Power On, Noninvasive Blood
Pressure performance, and all Safety tests
SpO
2
hardware upgrade M2703AU Option B72
Perform Visual Inspection, Power On, SpO
2
performance, and all Safety tests
Noninvasive Blood Pressure/SpO
2
hardware upgrade
M2703AU Option B73
Perform Visual Inspection, Power On, Noninvasive Blood
Pressure/SpO
2
performance and all Safety tests
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...