7 Testing and Maintenance
63
General Test Setup for the Performance of Safety Tests
Connect the detachable power cord of the device under test to the safety analyzer's test mains port.
For testing the applied part leakage current, connect all applied parts to the safety analyzer using the
appropriate patient lead or adapter cable. If necessary, repeat the safety test procedure until all available
applied parts have been tested. Refer to the documentation that accompanies the safety analyzer for
further details on how to set up and perform the test. Power on the device under test for all tests.
Protective Earth Resistance Test - Setup Example
Equipment Leakage Current Test - Setup Example
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...