7 Testing and Maintenance
86
Multi-Port Nurse Call Connector Test (Flexible Nurse Call)
This test checks the operation of the Flexible Nurse Call Relay. The Nurse Call Relay test is
recommended for customer sites where the nurse call is in use. The following diagram and table show
the pins and relay identifiers of the connector:
*1: The power loss indication functionality of the Nurse Call Relay board is not supported with fetal monitors.
The Nurse Call relay functions as follows:
•
During standard operation R1,R2,R3 _opener are closed; R1,R2,R3_closure are open.
•
During alarm condition—R1,R2,R3_opener are open; R1,R2,R3_closure are closed.
Tools required:
Ohmmeter.
1
Plug an M8087-61001 cable into the Nurse Call Relay connector.
2
Connect the ohmmeter and measure the pins as indicated in the diagram and table.
3
The relay contacts should behave as described above. The behavior may vary depending on
configuration choices. See the Configuration Guide for details on Alarm Relay settings.
Pin
Cable Color Coding
Relay
1
black
R2-closure
2
brown
R2-middle
3
red
R2-opener
4
orange
R3-closure
5
yellow
R3-middle
6
green
R3-opener
7
blue
n/a
8
purple
n/a
9
gray
n/a
10
white
n/a
11
pink
R1-closure
12
light green
R1-middle
13
black/white
R1-opener
14
brown/white
n/a
15
red/white
n/a
16
orange/white
n/a
17
blue/white
R_failure_closure
*1
18
purple/white
R_failure_middle
*1
19
green/white
R_failure_opener
*1
20
red/black
n/a
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......
Страница 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...