10 Repair and Disassembly
202
Removing the Front End Assembly
The front end assembly consists of:
•
the fetal sensor socket connectors with ribbon cable
•
the bus master board with ribbon cable
•
the optional noninvasive blood pressure and SpO
2
socket connectors
•
the connector block frame
To remove front end assembly:
1
Remove the top cover assembly (see “Removing the Top Cover Assembly” on page 141).
2
If the noninvasive blood pressure assembly is fitted, remove it (see “Removing the Noninvasive
Blood Pressure Assembly” on page 188).
3
If the SpO
2
assembly is fitted, remove it (see “Removing the Top Cover Assembly” on page 141).
Also remove the SpO
2
board holder by removing the two screws fastening it to the bottom
housing.
4
Disconnect the flat ribbon cable connecting the bus master board to the main CPU board.
Содержание Avalon FM20
Страница 2: ......
Страница 6: ...6 ...
Страница 38: ...3 Installation Instructions 38 ...
Страница 46: ...4 Theory of Operation 46 ...
Страница 145: ...10 Repair and Disassembly 145 7 The top cover assembly is now separated from the bottom housing assembly ...
Страница 154: ...10 Repair and Disassembly 154 10 Take the new rear housing cover and place it in the position shown ...
Страница 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Страница 163: ...10 Repair and Disassembly 163 4 Turn over the top cover assembly and place it top down on a cloth or other soft surface ...
Страница 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Страница 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Страница 200: ...10 Repair and Disassembly 200 2 Remove the screws holding the metal shield 3 Remove the shield ...
Страница 254: ...12 Specifications 254 ...
Страница 259: ......