10 Repair and Disassembly
192
Replacing the Noninvasive Blood Pressure Assembly
1
The new noninvasive blood pressure assembly kit comes with an insulator. Place the insulator on
the metal holder as shown.
2
Place the noninvasive blood pressure assembly onto the metal holder, ensuring the insulator is in
place.
3
Refit the three screws securing the NiBP module to the holder.