68 . Maintenance
4.
Remove the detector compartment cover by carefully lifting vertically using the two
hand holds, as shown in Figure 27.
Figure 27 Removing the detector compartment cover
5.
Move the detector unit carefully to the right and remove from the spectrometer.
The detector unit is installed by performing this procedure in reverse, tightening the
retaining screws and making sure the blanking caps are in place.
WARNING
Make sure you screw down the detector unit as this creates the earth
bonding for elctrical safety.
Details of the installation procedure for a sphere or Universal Reflectance Accessory (URA)
are given in the relevant manual on the
Lambda 650/850/950 User Manuals CD.