60 . Maintenance
Replacing a Lamp
If a lamp failure warning message is displayed, or you are seeing excessive noise on the
baseline, you can replace the lamps as follows.
WARNING
Electrical Hazard
High voltages are present at the lamp connectors in the lamp
compartment – severe electric shock hazard.
Never unplug or plug in a lamp cable while power is ON.
Switch off the spectrometer and remove the plug from the electrical
supply before starting with the replacement.
High Temperatures - Risk of Burns
Lamps soon become very hot, make sure that the lamps and the lamp
compartment have cooled to room temperature before you touch them.
UV Radiation
The lamps emit intense UV radiation which can damage your eyes.
Do not open the lamp compartment when the lamps are on.
Do not gaze into a lighted lamp.
The source lamps are located in the lamp compartment at the top left rear of the
spectrometer, as shown in Figure 18.
Figure 18 Lamp compartment
Lamp compartment
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...