![PerkinElmer Lambda 650 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/perkinelmer/lambda-650/lambda-650_hardware-manual_1530215047.webp)
Lambda 650/850/950 Hardware Guide . 47
The Single Cell Holder
Description
There are two single-cell holders provided with the instrument, one for the sample beam and
one for the reference beam. The single-cell holders are mounted on a plinth to bring them in
line with the radiation beam.
Figure 12 Single Cell Holder B0505071
Installing the Cell Holder
Install the plinth in the sample compartment as follows:
1.
Lower the plinth so that the two locating holes slip onto the two locating pins on the
baseplate in the sample compartment, as shown in Figure 13.
The plinth only fits one way round.
Figure 13 Fitting the plinth to the baseplate in the sample compartment
2.
Screw the two thumbscrews by hand into the baseplate.
Install the single-cell holder in the sample compartment as follows:
3.
Orientate the holder so that the word
LAMBDA
is toward the front of the sample
compartment (see Figure 14 below).
4.
Lower the holder so that the two locating holes slip onto the two studs on the two
locating pins on the plinth in the sample compartment.
Locating Pin
Thumbscrew
Thumbscrew
Locating Pin
Locking screw
for horizontal
alignment
Vertical adjustment
screw
Milled post
Lifter
Locking screw
for horizontal
alignment
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...