42 . Lambda 650 Installation
Calibrating the Instrument
Once the instrument is set up in the software you will need to calibrate it.
NOTE:
Make sure that the instrument has been switched on for at least an hour before calibrating it.
14.
From the Instruments page of the Explorer, right-click on the new instrument and select
Calibrate
.
The Calibration Utility dialog is displayed.
The dialog displays the names of the routines, a description, the last calibration date of
the routine, and whether the test is enabled.
15.
To view (and change) the settings for a particular routine, highlight the routine in the
table and then click
Settings
.
The settings dialog for the selected routine is displayed.
NOTE:
Settings is not available for Slit Calibration.
16.
Select the routines to be performed by the calibration.
There are three calibration routines available -
UV/Vis Two Peak
(see page 42 for
details),
Slit Calibration
, and
0%T
(see page 43 for details). All of these routines
should be performed.
A check mark indicates the routine is selected and will be performed.
17.
Click
Calibrate
.
The calibrations are performed in the order listed in the table. Messages below the table
show the progress of the calibration.
When a calibration has passed, a green check mark is displayed in the first column of
the table. If the test fails, a red cross is displayed. In such an event please contact your
local PerkinElmer Service Representative.
18.
When the calibration has finished, click
Close
to close the dialog.
The date and time of the calibration is recorded in the instrument event log.
NOTE:
It is not possible to print the calibration results.
Your instrument is now ready to be used.
UV/Vis Two Peak Calibration
This calibrates the UV/Vis wavelength range.
¾
Select
Manual set
,
Manual set 1 peak
,
Manual set 2 peaks
,
Auto search 1 peak
,
or
Auto search 2 peaks
.
•
Manual set
enables you to enter the
offset
and
factor
(without the need to
determine the values in advance).
•
Manual set 1 peak
enables you to change the offset. A well known peak can be
shifted to an exact wavelength by entering values of
Old Peak 1
and
New Peak
1
.
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...