28 . System Description
Minimum Performance Specifications
Specification
Lambda 950
Lambda 850
Lambda 650
Principle
Double beam, double
monochromator, ratio recording
UV/Vis/NIR spectrophotometer
with microcomputer
electronics, controlled by DELL
PC or compatible personal
computer.
Double beam, double
monochromator, ratio
recording UV/Vis
spectrophotometer with
microcomputer electronics,
controlled by DELL PC or
compatible personal computer.
Double beam, double
monochromator, ratio
recording UV/Vis
spectrophotometer with
microcomputer electronics,
controlled by DELL PC or
compatible personal
computer.
Optical system
All reflecting optical system
(SiO
2
coated) with holographic
grating monochromator with
1440 Lines/mm UV/Vis blazed
at 240 nm and 360 Lines/mm
NIR blazed at 1100 nm, Littrow
mounting, sample thickness
compensated detector.
All reflecting optical system
(SiO
2
coated) with holographic
grating monochromator with
1440 Lines/mm UV/Vis blazed
at 240 nm, Littrow mounting,
sample thickness compensated
detector optics.
All reflecting optical system
(SiO2 coated) with
holographic grating
monochromator with
1440 Lines/mm UV/Vis
blazed at 240 nm, Littrow
mounting, sample thickness
compensated detector
optics.
Beam Splitting
System
Chopper (46+ Hz, Cycle:
Dark/Sample/Dark/ Reference,
Chopper Segment Signal
Correction CSSC).
Chopper (46+ Hz, Cycle:
Dark/Sample/Dark/ Reference,
Chopper Segment Signal
Correction CSSC).
Chopper (46+ Hz, Cycle:
Dark/Sample/Dark/
Reference, Chopper
Segment Signal Correction
CSSC).
Detector
Photomultiplier R6872 for high
energy in the whole UV/Vis
wavelength range.
Peltier cooled PbS detector for
NIR.
Photomultiplier R6872 for high
energy in the whole UV/Vis
wavelength range.
R955 Photomultiplier, giving
high energy throughout the
whole UV/Vis range.
Source Pre-aligned
tungsten-halogen
and deuterium.
Pre-aligned tungsten-halogen
and deuterium
Pre-aligned tungsten-halogen
and deuterium.
Wavelength Range
(N
2
purge required
below 185 nm)
175 nm–3300 nm
175 nm–900 nm
190 nm– 900 nm
UV/Vis Resolution
≤
0.05 nm
≤
0.05 nm
≤
0.17 nm
NIR Resolution
≤
0.20 nm
N/A
N/A
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...