Electrical Safety
The instrument has been designed to protect the operator from potential electrical hazards.
This section describes some recommended electrical safety practices.
WARNING
Electrical Hazard
Any interruption of the protective conductor inside or outside the
instrument or disconnection of the protective conductor (earth/ground)
terminal is likely to make the instrument dangerous.
Intentional interruption is prohibited.
Lethal voltages are present in the instrument
•
Even with the power switch OFF, line power voltages can still be
present within the instrument.
•
When the instrument is connected to line power, terminals may be
live, and opening covers or removing parts (except those to which
access can be gained without the use of a tool) is likely to expose
live parts.
•
Capacitors inside the instrument may still be charged even if the
instrument has been disconnected from all voltage sources.
Electrical Protection
Insulation: Class I as defined in IEC 61010-1.
Installation Category: The instruments are able to withstand transient overvoltage according
to Installation Category II as defined in IEC 61010-1 and IEC 664.
Pollution Degree: The equipment will operate safely in environments that contain non-
conductive foreign matter and condensation up to Pollution Degree 2 as defined in IEC
61010-1 and IEC 664.
Electrical Safety Issues
To ensure satisfactory and safe operation of the instrument, it is essential that the
green/yellow lead of the line power cord is connected to true electrical earth (ground).
If any part of the instrument is not installed by a PerkinElmer service representative, make
sure that the line power plug is wired correctly:
Cord Lead Colors
Terminal
International USA
Live Brown
Black
Neutral Blue White
Protective Conductor
(earth/ground)
Green/Yellow Green
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...