34 . Lambda 650 Installation
Overview
CAUTION
Only the Lambda 650 is designed to be customer installed.
For the Lambda 850 or 950 please arrange for installation with your
local PerkinElmer Service Representative.
Requirements
For maximum stability and minimum maintenance, observe the following requirements when
choosing where to site the instrument:
•
A firm base free from vibration.
•
Enough space around and underneath the instrument for efficient air circulation, as
shown in Figure 8.
•
A constant temperature between 15 ºC and 35 ºC.
•
Constant humidity between 20% and 80% relative humidity.
•
An atmosphere free from dust and corrosive fumes.
•
Keep out of direct sunlight. Illumination with diffuse lighting is ideal.
•
A suitable source of electrical power should be located in the vicinity of the instrument,
in this case a proper earth-grounded 3-wire electrical outlet.
•
The standard sample compartment baseplates have drain holes in them to run off
spilled liquids to the benchtop underneath the instrument. If required, place a sheet of
thick filter paper under the instrument.
Figure 8 Space Requirements
74 cm (30 inches)
102 cm (41 inches)
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...