Lambda 650/850/950 Hardware Guide . 21
Fuses
WARNING
Electrical hazard
To prevent potential injury to yourself and damage to the instrument,
switch OFF all instruments in the system and disconnect them from the
line power supply before you alter, or make any new, electrical
connections.
The fuses are located in a fuse holder at the rear of the instrument, as shown in Figure 3.
Figure 3 Rear view, fuse holder
1.
Switch off the instrument and remove the line power cord from the electrical supply.
2.
Squeeze in and lever out the two lugs at each side of the fuse holder one at a time
using a flat-headed screwdriver
3.
Gently pull out the fuse holder.
Figure 4 Removing the fuse holder
4.
Replace the fuses with new ones of the same type and rating (3.15 A time delay).
NOTE:
The module has two fuses and you should always replace both at the same time, even if only
one blew, as the other may have been weakened.
5.
Replace the fuse holder by aligning the lug at the bottom of the fuse holder with the
slot in the socket and then pushing the holder into place.
A click is heard as each lug snaps into place.
NOTE:
If you use the correct fuses but the instrument still does not work correctly, or the fuses
blow repeatedly, contact your PerkinElmer office or representative.
Fuse Holder
Fuse Holder
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...