Lambda 650/850/950 Hardware Guide . 25
Optical System
The Lambda 650/850/950 Spectrometer features an all-reflecting, double-monochromator
optical system. The optical components are coated with silica for durability. Holographic
gratings are used in each monochromator for the UV/Vis range and the NIR range.
The optical system is depicted schematically in Figure 7.
Figure 7 Schematic of Optical System
Two radiation sources, a deuterium lamp (DL) and a halogen lamp (HL), cover the working
wavelength range of the spectrometer.
For operation in the near infrared (NIR) and visible (Vis) ranges, source mirror M1 reflects
the radiation from the halogen lamp onto mirror M2. At the same time it blocks the radiation
from the deuterium lamp.
M6
SBA
Dp
Detector
Compartment
M11
M11
΄
PM
(below PbS)
M8
Sample
Compartment
M13
CBM
M9
M10
M10
΄
C
RBA
DL
HL
Monochromator I
M2
M1
M3
G2
NIR
Monochromator II
FW
UV/Vis
G1
M5
M7
SA
M4
NIR
UV/Vis
S
R
M12
M12
΄
M13
΄
M14
PbS
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...