![PerkinElmer Lambda 650 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/perkinelmer/lambda-650/lambda-650_hardware-manual_1530215049.webp)
Lambda 650/850/950 Hardware Guide . 49
7.
By holding a piece of matt white paper behind each cell holder, visually examine the
light spot to see that the radiation beam is passing through the cell sample area.
Diffraction patterns become apparent if the radiation beam impinges on the cell wall.
8.
If the radiation beam is not centered exactly, loosen the two locking screws and the two
milled posts on the relevant cell holder and shift the cell holder plate to center the
radiation beam.
Then retighten the two milled posts and the two locking screws.
9.
Visually check the vertical alignment of the radiation beam in the cell sample area.
Alignment is correct when the radiation beam is just above the floor of the cell sample
area (minimum 2 mm) or covers the cell window.
NOTE:
The center of the window for micro flowcells should be ideally approximately 15 mm above
the base of the cell.
Figure 15 Correct alignment of the radiation beam in the cell sample area
10.
If alignment is required, turn the vertical adjustment screw on the lifter either clockwise
to raise the cell, or counterclockwise to lower the cell.
11.
Recheck the horizontal alignment of the radiation beam through the cell and correct if
necessary.
12.
Slew the monochromator to any value above 200 nm.
13.
Remove the card blocking the sample beam window and close the sample compartment
cover.
This completes the coarse alignment of the cell holders. If necessary, proceed with the
fine alignment as described below.
Fine Alignment
If fine alignment is necessary, proceed as follows:
1.
Slew the monochromator to your measurement wavelength or to 460 nm.
2.
Call up a method that uses transmission (%T) as the ordinate.
3.
If necessary change the ordinate mode to transmission.
4.
Remove the reference cell from the sample compartment.
Min. 2 mm
Содержание Lambda 650
Страница 1: ...LAMBDA 650 850 950 Hardware Guide UV VIS SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Introduction ...
Страница 11: ...Safety Practices ...
Страница 23: ...System Description ...
Страница 32: ...32 System Description ...
Страница 33: ...Lambda 650 Installation ...
Страница 44: ...44 Lambda 650 Installation ...
Страница 45: ...Using the Instrument ...
Страница 54: ...54 Using the Instrument ...
Страница 55: ...Maintenance ...
Страница 72: ...72 Maintenance ...