Measurement Settings
78
ZG User’s Manual
3
Adjust the bottom right of the
desired measurement area by the
↑
,
↓
,
←
and
→
keys, and press the
SET key.
The measurement region is set to
the specified area.
Profile
Noise filtering, output at measurement failure, and other options can be set in more detail.
Adjust the conditions when the measurement cannot be performed properly.
Setting of profiles is enabled only in the EXP mode.
Setting the average number of times
Changes in data are smoothed out using the average values of adjacent data. Smoothing
is performed in the “width direction (X-axis direction).”
FUN mode-[MEAS]-[IMAGE]-[PROFILE]-[AVERAGE]
Setting value
Description
1, 2, 4, 8, 16, 32, 64
Sets the number of data to average.
(default value: 4 when the CCD mode is NORMAL or HI-RESO, and
1 when the CCD mode is HI-SPEED)
↑↓