Checking/Adjusting the Measurement Status - ADJ Mode
51
ZG User’s Manual
2
BAS
IC OPERATIONS
Setting the Judgment Value
Setting the Judgment Threshold Value
Set the range of measurement values to be judged as OK.
ADJ mode-MENU/VIEW key-[JUDGE]
The default judgment threshold value is the “rated measurement range of the currently
connected Sensor
÷
4”.
Example:As the measurement range of the ZG-WDS70 (diffuse reflection type) is “60 mm” (±30
mm), the default judgment threshold value becomes “60
÷
4=15”. So, the default
threshold judgment value is ±15 mm.
Setting the Zero Reset Offset Value
To set a reference value for zero reset to a value other than 0 (zero), set the offset amount
using this function. After setting any target value, execute a zero reset in the RUN mode.
Setting Zero Reset p.56
ADJ mode-MENU/VIEW key-[ZERO]
Setting value
Description
HIGH
Sets the HIGH threshold value.
Range: -999.99999 to +999.99999
LOW
Sets the LOW threshold value.
Range: -999.99999 to +999.99999
To change numerical values, use the
↑
UP key/
↓
DOWN key, and to change the number of
digits use the
←
LEFT key/
→
RIGHT key.
Setting value
Description
Zero
Sets the offset amount.
Range: -999.99999 to +999.99999 (default value: 0)
To change numerical values, use the
↑
UP key/
↓
DOWN key, and to change the number of
digits use the
←
LEFT key/
→
RIGHT key.
N
ot
e
Important
N
ot
e
Содержание ZG -
Страница 12: ...10 ZG User s Manual MEMO ...
Страница 44: ...Overview of Settings and Measurement 42 ZG User s Manual MEMO ...
Страница 60: ...Functions Operations Used during Operation RUN Mode 58 ZG User s Manual MEMO ...
Страница 110: ...System Settings 108 ZG User s Manual MEMO ...
Страница 156: ...Serial Communication 154 ZG User s Manual MEMO ...
Страница 206: ...204 ZG User s Manual MEMO ...
Страница 207: ...205 ZG User s Manual 5 APPENDICES MEMO ...
Страница 208: ...206 ZG User s Manual MEMO ...