Measurement Settings
71
ZG User’s Manual
3
FUNCTION SETTINGS
Setting Scaling (Correction Processing)
If scaling is set, differences between measurement values and actual sizes, that occur
due to the color, material or other factors of the measurement target, can be corrected.
There are two scaling setup modes, “automatic setting” and “manual setting.” In the
automatic setting mode, actual measurement is performed, and in the manual setting
mode, the correction values are set manually.
(default value: OFF)
The settings below return to the default settings when scaling is set. Set these items after
scaling settings have been completed.
• Zero reset
Measurement items
Setup Method
Reference
2-pt step, 3-pt step, edge
width
Automatic setting: one-point scaling
Measurement is performed once, and the correction
value for the measurement value is set.
p.72
Height, edge position
Automatic setting: two-point scaling
Measurement is performed at two positions, and
correction values are set for those measurement
values.
p.72
Angle, cross-sectional
area
Manual setting
Input and set coefficients and offset values directly as
numerical values.
p.73
5 mm
5 mm
4.95 mm
Measurement value
This error is corrected.
Actual size value
Important
Содержание ZG -
Страница 12: ...10 ZG User s Manual MEMO ...
Страница 44: ...Overview of Settings and Measurement 42 ZG User s Manual MEMO ...
Страница 60: ...Functions Operations Used during Operation RUN Mode 58 ZG User s Manual MEMO ...
Страница 110: ...System Settings 108 ZG User s Manual MEMO ...
Страница 156: ...Serial Communication 154 ZG User s Manual MEMO ...
Страница 206: ...204 ZG User s Manual MEMO ...
Страница 207: ...205 ZG User s Manual 5 APPENDICES MEMO ...
Страница 208: ...206 ZG User s Manual MEMO ...