I/O Timing Charts
124
ZG User’s Manual
External input of bank switching signal
External input of laser stop signal
T1:
ENABLE signal OFF
response time
This is the time after input of the back switching signal until the
ENABLE signal turns OFF.
When the ECO monitor (display OFF) is operating: 20 ms or less
When the digital monitor is operating: 300 ms or less
When the profile monitor is operating: 200 ms or less
T2:
Bank switching time
This is the time in which bank switching is executed.
When the ECO monitor (display OFF) is operating: 200 ms or less
When the digital monitor is operating: 500 ms or less
When the profile monitor is operating: 400 ms or less
T1:
Laser stop response time
This is the time after the laser stop signal is input until laser
emission is stopped. 30 ms or less
T2:
Laser restore response
time
This is the time after the laser stop signal is canceled until laser
emission is started. 20 ms or less
T3:
ENABLE signal OFF
response time
This is the time after input of the laser stop signal until the ENABLE
signal turns ON. 10 ms or less
T4:
ENABLE signal ON
response time
This is the time after cancellation of the laser stop signal until the
ENABLE signal turns ON. 5 ms or less
T1
T2
Ban
k
switching signal
ON
OFF
ENABLE signal
OFF
ON
L
aser
stop
s
i
gna
l
L
aser
beam
em
i
ss
i
on
O
FF
O
N
O
N
O
FF
T
3
T
2
EN
AB
LE
s
i
gna
l
O
FF
O
N
T
1
T4
Содержание ZG -
Страница 12: ...10 ZG User s Manual MEMO ...
Страница 44: ...Overview of Settings and Measurement 42 ZG User s Manual MEMO ...
Страница 60: ...Functions Operations Used during Operation RUN Mode 58 ZG User s Manual MEMO ...
Страница 110: ...System Settings 108 ZG User s Manual MEMO ...
Страница 156: ...Serial Communication 154 ZG User s Manual MEMO ...
Страница 206: ...204 ZG User s Manual MEMO ...
Страница 207: ...205 ZG User s Manual 5 APPENDICES MEMO ...
Страница 208: ...206 ZG User s Manual MEMO ...