I/O Settings
94
ZG User’s Manual
3
Input the correction value
corresponding to output MIN, and
press the SET key.
4
Repeat steps 1 to 3 to set output
MAX.
Setting the Clamp Level
A preset clamp value (abnormal value) can be output in cases where the light density is
temporarily excessive or insufficient (i.e. measurement is not possible) due to defects or
holes, for example, in the measurement target.
FUN mode-[I/O]-[ANALOG]-[CLAMP]
To change numerical values, use the
↑
UP key/
↓
DOWN key, and to
change the number of digits use the
←
LEFT key/
→
RIGHT key.
Setting value
At current output: MIN (approx.2 mA), MAX (approx.25 mA, default value), 4 to 20 mA (in 1
mA increments)
At voltage output: MIN (approx.-11 V), MAX (approx.11 V, default value), -10 to 10 V (in 1 V
increments)