Measurement Settings
73
ZG User’s Manual
3
FUNCTION SETTINGS
Manual Setting Method
Span and offset can be set by inputting numerical values manually to fine-tune the
measurement values. These can be set for each individual task. Span and offset are
automatically set after scaling is executed. So, modify these settings as necessary.
Manual setting is enabled only in the EXP mode.
FUN mode-[MEAS]-[SENS]-[TASK1 to 8]-[SCALE]-[MANUAL]
Setting value
Description
SPAN
Sets the inclination of the sensor characteristics as a coefficient.
Range: -4.0000 to 4.0000
OFFSET
Adds/subtracts a fixed value to or from the measured value.
Range: -999.99999 to 999.99999
-4.
0000
(mm)
4.
0000
Measurement value
Displacement of
measurement target
(mm)
Measurement value
Displacement of
measurement target