Measurement Settings
81
ZG User’s Manual
3
FUNCTION SETTINGS
Setting the number of interpolated pixels
Set the number of pixels to be interpolated when interpolating profile data. Profile data is
interpolated only if missing areas (areas where measurement data cannot be obtained)
contain less than the specified number of pixels. Interpolation is not performed if a
number of pixels greater than the specified number cannot be measured continuously.
This feature can be applied, for example, to the measurement of measurement targets
with holes in them.
Example: When the number of interpolated pixels is set to 2
FUN mode-[MEAS]-[IMAGE]-[PROFILE]-[SUPnum]
Setting value
Description
1, 2, 4, 8, 16, ALL (unit:
pixels)
Interpolation is performed when missing areas contain less than
the specified number of pixels.
When [ALL] is selected, interpolation is performed on the entire
profile regardless of the number of pixels.
(default value: 4)
Areas where measurement
values could not be obtained
Height direction
Width direction
Not interpolated
(3 pixels > 2 pixels)
Interpolated
(1 pixel
≤
2 pixels)