Auto Focus Multi Code Reader V430-series User’s Manual (Z392)
11-11
Configuration Database
Image Process Looping
Disabled
When
Image Process Looping
is set to
Disabled
, an image is captured for every database
configuration, whether or not any camera settings have been modified.
Enabled
When
Image Process Looping
is set to
Enabled
, the last captured image frame is re-processed
with the new IP and decode parameters. If camera settings have not been changed from
the last capture event, and when a database configuration is loaded to current, active
settings, no new image is captured.
Note:
An exception to this is the first database index: a new image is always captured
when the first database index is used. When a camera setting has been modified from one
database setting to the next, a new image is captured. For example, if all database entries
contained the same camera setting values but had different IP and decode parameters, an
image frame would only be captured when the first database configuration was used.
Usage:
Useful in applications where it is necessary to process a single captured
image multiple times using different IP and decode parameters.
Serial Cmd:
<
K256,
switch mode,frame count/time,
image process looping
,image dimensions
>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Содержание V430-F-series
Страница 1: ...Auto Focus Multi Code Reader V430 F series User s Manual Z392 E1 02 ...
Страница 9: ...CONTENTS 7 Auto Focus Multi Code Reader V430 series User s Manual Z392 ...
Страница 97: ...4 14 Auto Focus Multi Code Reader V430 series User s Manual Z392 Additional Notes about Calibration ...
Страница 125: ...5 28 Auto Focus Multi Code Reader V430 series User s Manual Z392 Good Decode s Needed ...
Страница 265: ...8 30 Auto Focus Multi Code Reader V430 series User s Manual Z392 ISO IEC 29158 Symbol Quality Output ...
Страница 279: ...9 14 Auto Focus Multi Code Reader V430 series User s Manual Z392 New Master Pin ...
Страница 325: ...11 18 Auto Focus Multi Code Reader V430 series User s Manual Z392 Request All Configuration Database Settings ...
Страница 403: ...Omron Microscan Systems Inc 700 SW 39th St Suite 100 Renton WA 98057 USA ...