5-20
Auto Focus Multi Code Reader V430-series User’s Manual (Z392)
Capture Mode
Triggered Capture
Usage:
Useful in applications where each decode must be treated as a discrete
event, regardless of symbol data.
Definition:
The first trigger event starts the read cycle, and subsequent triggers will
continue until the predetermined
Number of Captures
is met, or until the
predetermined
End of Read Cycle
condition is met-- whichever occurs first.
Note:
If
End of Read Cycle
is set for
New Trigger
and the read cycle
qualifications have not been met, the read cycle will only end
once it
receives the first trigger after reaching the predetermined
Number of
Captures
setting
.
Processing
Read cycle ends
on New Trigger
or Timeout, as
configured.
Captures on every trigger
Capture
First trigger
starts the read
cycle
Содержание V430-F-series
Страница 1: ...Auto Focus Multi Code Reader V430 F series User s Manual Z392 E1 02 ...
Страница 9: ...CONTENTS 7 Auto Focus Multi Code Reader V430 series User s Manual Z392 ...
Страница 97: ...4 14 Auto Focus Multi Code Reader V430 series User s Manual Z392 Additional Notes about Calibration ...
Страница 125: ...5 28 Auto Focus Multi Code Reader V430 series User s Manual Z392 Good Decode s Needed ...
Страница 265: ...8 30 Auto Focus Multi Code Reader V430 series User s Manual Z392 ISO IEC 29158 Symbol Quality Output ...
Страница 279: ...9 14 Auto Focus Multi Code Reader V430 series User s Manual Z392 New Master Pin ...
Страница 325: ...11 18 Auto Focus Multi Code Reader V430 series User s Manual Z392 Request All Configuration Database Settings ...
Страница 403: ...Omron Microscan Systems Inc 700 SW 39th St Suite 100 Renton WA 98057 USA ...