EX Series Operating Manual
*RCL {1|2|3|4|5|6|7|8|10}
This command is recalling the power supply's saved data in the non-volatile memory
You can select among 1 ~ 10 number.
> 1 ~ 10
Memory sector.
ex) *rcl 2
Recall and apply the number 2 memory data to the power supply.
*CLS
Clear the all error event.
ex) *CLS
*RST
This command is initializing of the power supply. Following is initializing contents.
This command is initializing of the power supply. Following is initializing contents.
> Output
OFF
> Voltage Limit
0V
> Current Limit
Maximum current value of the power supply.
> OVP
Maximum OVP value of the power supply.
> OCP
Maximum OCP value of the power supply.
> UVL
0V
> OVL
Voltage Limit MAX value.
> UCL
0A
> OCL
Current Limit MAX value.
> Volt:Step
Minumum value of setting
> Curr:Step
Minumum value of setting
> Key Lock
OFF
> Factory Mode
Keep the value of before reset
y
p
> Trip상태
Clear if the OVP or OCP Tripped occurred automatically.
ex) *RST
Initialize the power supply.
- 98 -
Содержание EX Series
Страница 2: ......
Страница 104: ...EX Series Operating Manual 104...
Страница 105: ...EX Series Operating Manual 105...
Страница 107: ...EX Series Operating Manual 2U 2 6KW 5KW Product Outside Dimension 2U 2 6KW 5KW Product Rear Feature 107...
Страница 108: ...EX Series Operating Manual 4U 5 1KW 10KW Product Outside Dimension 4U 5 1KW 10KW Product Rear Feature 108...
Страница 113: ...EX Series Operating Manual Reliability Durability Life time Test System Solution 113...