EX Series Operating Manual
8-6. Interface Commands Error
This chapter is for PC interface communicaton error.
-120, "Suffix too long"
A suffix for a numeric parameter contained too long characters. Maximum memory buffer
that you are able to send is 50 byte at only once
that you are able to send is 50 byte at only once.
-121, "Invalid data"
An invalid character was found in the command.
ex) volt 10V
Don't need character 'V'.
You must revise to "volt 10".
-122, "Syntax error"
Invalid syntax was found in the command string. You may have inserted a blank space
Invalid syntax was found in the command string. You may have inserted a blank space
before or after a colon in the command header, or before a comma.
ex)volt
You skipped the numeric value.
You must revise to "volt 10".
-123, "Invalid suffix"
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the
suffix.
ex)volt 10*
You added "*" after command
ex)volt 10*
You added "*" after command
You must revise to "volt 10".
-124, "Undefined header"
A command was received that is not valid for this power supply. You may have
misspelled the command or it may not be a valid command. If you are using the short
form of the command, remember that it may contain up to four letters.
ex)volta 10
The power supply just receive "volt" or "voltage".
You must revise to "volt 10".
-220,"No execution"
The power supply does not accept to execute.
ex)Volt 20
Set the voltage to 20V.
Volt:OVP 15
The error occurred when you set the OVP level lower
than setting voltage value.
221 "Setting conflict"
-221, "Setting conflict"
Indicates that a legal program data element was parsed but could not be executed due to
the current device state.
ex)POL N
This is polarity change command, but could not be executed
in current device.
-222, "Out of data"
A numeric parameter value is outside the valid range for the command.
ex)volt 1000
Voltage setting value is too much big.
You must revise to "volt 10".
-223, "Incorret error"
It appears when you send new query command again without receiving former query
command execution from buffer.
ex)*idn?
You sent the query command to the power supply and
volt?
immidiately send again another new command.
volt?
immidiately send again another new command.
- 102 -
Содержание EX Series
Страница 2: ......
Страница 104: ...EX Series Operating Manual 104...
Страница 105: ...EX Series Operating Manual 105...
Страница 107: ...EX Series Operating Manual 2U 2 6KW 5KW Product Outside Dimension 2U 2 6KW 5KW Product Rear Feature 107...
Страница 108: ...EX Series Operating Manual 4U 5 1KW 10KW Product Outside Dimension 4U 5 1KW 10KW Product Rear Feature 108...
Страница 113: ...EX Series Operating Manual Reliability Durability Life time Test System Solution 113...