EX Series Operating Manual
≫ Related Remote Interface Command
REM:VPROG {LOC|REM}
REM:VPROG?
Application:
REM:VPROG?
Return Value : "LOC"
REM:VPROG REM
Setting Remote Mode
Remote Current Programming
By using 0~5VDC input, you can program the output current from 0V to 100% current
The following is a pin connection drawing.
< Figure 4-4 Remote Current Programming >
< Figure. 4 4 Remote Current Programming >
Remote Current Programming setting
The factory default is set as local mode. If device is equipped with option, you should change
the setting and you can use remote current programming.
Warning
Before setting, you must complete external wiring.
If you do external wiring under the remote cotrolling, the DUT can be damaged because of power
supply's output current floating.
▌Turn on the power supply
After turned on, check the power supply displays"
**OUTPUT OFF**
" Mode.
▌Press the limit display key and hold 3 seconds.
LCD Display
1. V PROG
▌Press the cursor key to move to "I Prog" and change to "I PROG".
LCD Display
2. I PROG
▌In other to use remote current programming, press limit display key.
LCD Display
local
- 57 -
Содержание EX Series
Страница 2: ......
Страница 104: ...EX Series Operating Manual 104...
Страница 105: ...EX Series Operating Manual 105...
Страница 107: ...EX Series Operating Manual 2U 2 6KW 5KW Product Outside Dimension 2U 2 6KW 5KW Product Rear Feature 107...
Страница 108: ...EX Series Operating Manual 4U 5 1KW 10KW Product Outside Dimension 4U 5 1KW 10KW Product Rear Feature 108...
Страница 113: ...EX Series Operating Manual Reliability Durability Life time Test System Solution 113...