10.6-3
Fig. 10.6-2 BLD cover
Fig. 10.6-3 Front view of BLD TYPE II
10.6.2. TYPE II: RK-GR2
10.6.2.1. Zero revision of blood leak detector
Adjust the zero revision of blood leak detector when “6.BLD AUTO ZERO” value less than
-150ppm in maintenance mode 1 screen during preparation process.
(1) Press [Prep.] SW to shift to preparation process.
(2) Remove BLD cover of rear side of machine as shown in Fig.10.6-2.
(3) Press
inside blood leak detector as shown in Fig.10.6-3 for 1 second to 3second after
preparation time pass over 5 minutes.
(4) Check green LED inside blood leak detector blinking twice as shown in Fig.10.6-3.
(5) Check “6. BLD AUTO ZERO” value more –80ppm in maintenance mode 1 screen.
ܭ
CAUTION
z
Perform acid and chemical rinse before you calibrate it.
z
Check no bubble inside cartridge of blood leak detector when zero revision is performed.
z
Can not revise if pressed SW time is less than 1 second. (Green LED also do not blink.)
z
Initialize zero revision value if pressed SW time is more than 5 second.
Green LED blinks twice.
Содержание SURDIAL
Страница 2: ......
Страница 4: ...PREFACE 2...
Страница 16: ...CONTENTS 8...
Страница 18: ...1 2...
Страница 19: ...2 1 2 CONTRAINDICATION Any application other than dialysis using SURDIAL is contraindicated...
Страница 20: ...2 2...
Страница 22: ...3 2...
Страница 24: ...4 2...
Страница 25: ...5 FLOW CHART...
Страница 26: ......
Страница 32: ...5 6...
Страница 33: ...6 DISASSEMBLING AND REPLACEMENT...
Страница 34: ......
Страница 40: ...6 1 1 6...
Страница 44: ...6 1 2 4...
Страница 48: ...6 1 3 4...
Страница 50: ...6 1 4 2...
Страница 54: ...6 1 6 2...
Страница 56: ...6 2 1 2...
Страница 60: ...6 2 2 4...
Страница 64: ...6 2 3 4...
Страница 66: ...6 2 4 2...
Страница 72: ...6 2 5 6...
Страница 73: ...6 2 6 1 6 2 6 Blood pump Assy BP BPC Fig 6 2 6 1 Blood pump Assy...
Страница 76: ...6 2 6 4...
Страница 77: ...6 2 7 1 6 2 7 Gear pump Assy P1 P2 6 2 7 1 TYPE I Micro gear pump Fig 6 2 7 1 Micro gear pump Assy TYPE I...
Страница 79: ...6 2 7 3 6 2 7 2 TYPE II Magnet gear pump Normal Type Fig 6 2 7 2 Magnet gear pump Assy TYPE II Normal Type...
Страница 81: ...6 2 7 5 6 2 7 3 TYPE II Magnet gear pump A Type Fig 6 2 7 3 Magnet gear pump Assy TYPE II A Type...
Страница 84: ...6 2 7 8...
Страница 85: ...6 2 8 1 6 2 8 Venous clamp Assy CLV Fig 6 2 8 1 Venous clamp Assy...
Страница 87: ...6 2 9 1 6 2 9 Syringe pump Assy SP Fig 6 2 9 1 Syringe pump Assy...
Страница 92: ...6 3 2 2...
Страница 94: ...6 3 3 2...
Страница 96: ...6 3 4 2...
Страница 100: ...6 3 5 4...
Страница 106: ...6 3 7 4...
Страница 114: ...6 3 10 2...
Страница 115: ...6 3 11 1 6 3 11 Chamber Assy VCa VCb 6 3 11 1 Chamber Assy Overall Fig 6 3 11 1 Chamber Assy Overall Chamber...
Страница 117: ...6 3 11 3 6 3 11 2 Chamber Assy Viscous chamber Fig 6 3 11 2 Chamber Assy Viscous chamber Cross section A A...
Страница 132: ...6 3 16 2...
Страница 133: ...7 CONSTRUCTION OF ELECTRICAL PART...
Страница 134: ......
Страница 141: ...8 CONNECTION REPLACEMENT AND ADJUSTMENT...
Страница 142: ......
Страница 186: ...8 12 8...
Страница 195: ...9 MAINTENANCE MODES...
Страница 196: ......
Страница 198: ...9 1 2...
Страница 204: ...9 3 2...
Страница 220: ...9 9 6...
Страница 230: ...9 12 2...
Страница 232: ...9 13 2...
Страница 234: ......
Страница 246: ...10 3 6...
Страница 262: ...10 8 2...
Страница 264: ...10 9 2...
Страница 268: ...10 10 4...
Страница 270: ...10 11 2...
Страница 280: ...10 15 4...
Страница 290: ...10 17 2...
Страница 292: ...10 18 2...
Страница 298: ...10 21 2...
Страница 300: ...10 22 2...
Страница 303: ...11 DIP SWITCHES...
Страница 304: ......
Страница 306: ...11 1 2...
Страница 312: ...11 2 6...
Страница 313: ...12 VARIOUS DATA...
Страница 314: ......
Страница 316: ...12 1 2...
Страница 334: ...12 3 12...