10.6-2
(11) Repeat the above steps (9) and (10) until each of 0±55 and 800±55 is set.
Adjust R41 (rough adjustment) and R47 (fine adjustment) if each value can not be adjusted
only R47.
Fig. 10.6-1(b) Blood leak detector PCB
NOTE
z
R41 has large sensitivity because it is used for rough adjustment. Adjust R41 (rough adjustment)
and R47 (fine adjustment). Total value in both volume become adjustment values due to R41
connect to R47 in series. Adjustment value is not stable when adjusted point of R47 is shifted.
Adjusted point of R47 is required to return to center point when R41 adjust. The center point of
R47 is seventh rotation. (Total of R47 is fourteenth rotation.)
(12) Attach the removed sensor cover to the sensor to return it to the original position.
(13) Return the removed left cover to the original position.
Содержание SURDIAL
Страница 2: ......
Страница 4: ...PREFACE 2...
Страница 16: ...CONTENTS 8...
Страница 18: ...1 2...
Страница 19: ...2 1 2 CONTRAINDICATION Any application other than dialysis using SURDIAL is contraindicated...
Страница 20: ...2 2...
Страница 22: ...3 2...
Страница 24: ...4 2...
Страница 25: ...5 FLOW CHART...
Страница 26: ......
Страница 32: ...5 6...
Страница 33: ...6 DISASSEMBLING AND REPLACEMENT...
Страница 34: ......
Страница 40: ...6 1 1 6...
Страница 44: ...6 1 2 4...
Страница 48: ...6 1 3 4...
Страница 50: ...6 1 4 2...
Страница 54: ...6 1 6 2...
Страница 56: ...6 2 1 2...
Страница 60: ...6 2 2 4...
Страница 64: ...6 2 3 4...
Страница 66: ...6 2 4 2...
Страница 72: ...6 2 5 6...
Страница 73: ...6 2 6 1 6 2 6 Blood pump Assy BP BPC Fig 6 2 6 1 Blood pump Assy...
Страница 76: ...6 2 6 4...
Страница 77: ...6 2 7 1 6 2 7 Gear pump Assy P1 P2 6 2 7 1 TYPE I Micro gear pump Fig 6 2 7 1 Micro gear pump Assy TYPE I...
Страница 79: ...6 2 7 3 6 2 7 2 TYPE II Magnet gear pump Normal Type Fig 6 2 7 2 Magnet gear pump Assy TYPE II Normal Type...
Страница 81: ...6 2 7 5 6 2 7 3 TYPE II Magnet gear pump A Type Fig 6 2 7 3 Magnet gear pump Assy TYPE II A Type...
Страница 84: ...6 2 7 8...
Страница 85: ...6 2 8 1 6 2 8 Venous clamp Assy CLV Fig 6 2 8 1 Venous clamp Assy...
Страница 87: ...6 2 9 1 6 2 9 Syringe pump Assy SP Fig 6 2 9 1 Syringe pump Assy...
Страница 92: ...6 3 2 2...
Страница 94: ...6 3 3 2...
Страница 96: ...6 3 4 2...
Страница 100: ...6 3 5 4...
Страница 106: ...6 3 7 4...
Страница 114: ...6 3 10 2...
Страница 115: ...6 3 11 1 6 3 11 Chamber Assy VCa VCb 6 3 11 1 Chamber Assy Overall Fig 6 3 11 1 Chamber Assy Overall Chamber...
Страница 117: ...6 3 11 3 6 3 11 2 Chamber Assy Viscous chamber Fig 6 3 11 2 Chamber Assy Viscous chamber Cross section A A...
Страница 132: ...6 3 16 2...
Страница 133: ...7 CONSTRUCTION OF ELECTRICAL PART...
Страница 134: ......
Страница 141: ...8 CONNECTION REPLACEMENT AND ADJUSTMENT...
Страница 142: ......
Страница 186: ...8 12 8...
Страница 195: ...9 MAINTENANCE MODES...
Страница 196: ......
Страница 198: ...9 1 2...
Страница 204: ...9 3 2...
Страница 220: ...9 9 6...
Страница 230: ...9 12 2...
Страница 232: ...9 13 2...
Страница 234: ......
Страница 246: ...10 3 6...
Страница 262: ...10 8 2...
Страница 264: ...10 9 2...
Страница 268: ...10 10 4...
Страница 270: ...10 11 2...
Страница 280: ...10 15 4...
Страница 290: ...10 17 2...
Страница 292: ...10 18 2...
Страница 298: ...10 21 2...
Страница 300: ...10 22 2...
Страница 303: ...11 DIP SWITCHES...
Страница 304: ......
Страница 306: ...11 1 2...
Страница 312: ...11 2 6...
Страница 313: ...12 VARIOUS DATA...
Страница 314: ......
Страница 316: ...12 1 2...
Страница 334: ...12 3 12...