10.13-1
10.13. Adjustment of blood pump flow rate
10.13.1 In case of you use Standard Rolling Tube( 8× 12mm)
Adjust blood pump flow rate by setting blood line to blood pump to input GAIN and OFFSET values
of blood pump flow rate by referring to actually measured flow rate (discharged water weight or
volume/min) in maintenance mode 6 screen based on the following procedures.
(1) Required devices
Electronic balance or measuring cylinder
Stopwatch
(2) Turn on power SW of machine.
(3) Turn on [DIALYSIS] SW to shift to dialysis wait process.
(4) Release setting protection by selecting ON of “20.MAINTENANCE MODE FREE” in
maintenance mode 7 screen.
(5) Set blood line to blood pump.
(6) Place electronic balance or measuring cylinder below outflow end of the line to measure
discharged water weight or volume.
WATER
BP
Blood line
Electronic balance
Fig. 10.13-1 Adjustment of blood pump flow rate
(7) Set blood pump flow rate to 50mL/min and operate the pump.
(8) Measure the discharged water weight or volume per one minute.
Flow rate (g/min or mL/min) = measured weight or volume ÷ time (one minute)
(9) Input “18.BPF(OFFSET)” value of blood pump flow rate in maintenance mode 6 screen value
to try to adjust the actual flow rate to about 50mL/min. Decrease (increase) the value if the
actual flow rate is higher (lower) than the set flow rate.
Содержание SURDIAL
Страница 2: ......
Страница 4: ...PREFACE 2...
Страница 16: ...CONTENTS 8...
Страница 18: ...1 2...
Страница 19: ...2 1 2 CONTRAINDICATION Any application other than dialysis using SURDIAL is contraindicated...
Страница 20: ...2 2...
Страница 22: ...3 2...
Страница 24: ...4 2...
Страница 25: ...5 FLOW CHART...
Страница 26: ......
Страница 32: ...5 6...
Страница 33: ...6 DISASSEMBLING AND REPLACEMENT...
Страница 34: ......
Страница 40: ...6 1 1 6...
Страница 44: ...6 1 2 4...
Страница 48: ...6 1 3 4...
Страница 50: ...6 1 4 2...
Страница 54: ...6 1 6 2...
Страница 56: ...6 2 1 2...
Страница 60: ...6 2 2 4...
Страница 64: ...6 2 3 4...
Страница 66: ...6 2 4 2...
Страница 72: ...6 2 5 6...
Страница 73: ...6 2 6 1 6 2 6 Blood pump Assy BP BPC Fig 6 2 6 1 Blood pump Assy...
Страница 76: ...6 2 6 4...
Страница 77: ...6 2 7 1 6 2 7 Gear pump Assy P1 P2 6 2 7 1 TYPE I Micro gear pump Fig 6 2 7 1 Micro gear pump Assy TYPE I...
Страница 79: ...6 2 7 3 6 2 7 2 TYPE II Magnet gear pump Normal Type Fig 6 2 7 2 Magnet gear pump Assy TYPE II Normal Type...
Страница 81: ...6 2 7 5 6 2 7 3 TYPE II Magnet gear pump A Type Fig 6 2 7 3 Magnet gear pump Assy TYPE II A Type...
Страница 84: ...6 2 7 8...
Страница 85: ...6 2 8 1 6 2 8 Venous clamp Assy CLV Fig 6 2 8 1 Venous clamp Assy...
Страница 87: ...6 2 9 1 6 2 9 Syringe pump Assy SP Fig 6 2 9 1 Syringe pump Assy...
Страница 92: ...6 3 2 2...
Страница 94: ...6 3 3 2...
Страница 96: ...6 3 4 2...
Страница 100: ...6 3 5 4...
Страница 106: ...6 3 7 4...
Страница 114: ...6 3 10 2...
Страница 115: ...6 3 11 1 6 3 11 Chamber Assy VCa VCb 6 3 11 1 Chamber Assy Overall Fig 6 3 11 1 Chamber Assy Overall Chamber...
Страница 117: ...6 3 11 3 6 3 11 2 Chamber Assy Viscous chamber Fig 6 3 11 2 Chamber Assy Viscous chamber Cross section A A...
Страница 132: ...6 3 16 2...
Страница 133: ...7 CONSTRUCTION OF ELECTRICAL PART...
Страница 134: ......
Страница 141: ...8 CONNECTION REPLACEMENT AND ADJUSTMENT...
Страница 142: ......
Страница 186: ...8 12 8...
Страница 195: ...9 MAINTENANCE MODES...
Страница 196: ......
Страница 198: ...9 1 2...
Страница 204: ...9 3 2...
Страница 220: ...9 9 6...
Страница 230: ...9 12 2...
Страница 232: ...9 13 2...
Страница 234: ......
Страница 246: ...10 3 6...
Страница 262: ...10 8 2...
Страница 264: ...10 9 2...
Страница 268: ...10 10 4...
Страница 270: ...10 11 2...
Страница 280: ...10 15 4...
Страница 290: ...10 17 2...
Страница 292: ...10 18 2...
Страница 298: ...10 21 2...
Страница 300: ...10 22 2...
Страница 303: ...11 DIP SWITCHES...
Страница 304: ......
Страница 306: ...11 1 2...
Страница 312: ...11 2 6...
Страница 313: ...12 VARIOUS DATA...
Страница 314: ......
Страница 316: ...12 1 2...
Страница 334: ...12 3 12...