4.20 Turning the Output On or Off Rapidly
KP3000GS
165
4.20 Turning the Output On or Off Rapidly
In the factory default settings, the internal output relay of this product is activated linking to the
output on/off. In the output off state, the internal circuit, output terminal, and the outlet of this
product are physically separated. On the other hand, if the output relay chattering causes any
problem, or if you want to turn the output on or off rapidly, you can disable the activation of the
output relay. In that case, the output relay is fixed to the on state and the operation to turn the output
on/off is performed by the semiconductor device. The output off state means the state of high
impedance. Table 4-15 shows the list of the output terminal impedance (reference values in the rear
output terminal) in the output off state when the output relay control is disabled.
Table 4-15 The Output Terminal Impedance in the Output Off State when
the Activation of the Output Relay is Disabled
AC/DC mode
100 V range
200 V range
AC and ACDC
modes
200 k
Ω
4
N
+ 3
200 k
Ω
4
N
+ 3
200 k
Ω
2
N
+ 1
200 k
Ω
2
N
+ 1
DC mode
200 k
Ω
4
N
+ 1
200 k
Ω
4
N
+ 1
200 k
Ω
2
N
+ 1
200 k
Ω
2
N
+ 1
Note 1: The reference value of output terminal at rear panel.
Note 2:
N
indicates the number of the power units energized.
--------
Notes
------------------------------------------------------------------------------------------------------------
Even when the activation of the output relay is disabled, the relay is turned off if the protection
function is activated. While the error message appearing when the protection function is
activated displays "Press Enter Key," pressing the ENTER key clears the message and turns on
the output relay again although the output remains off.
The output relay is in the off state during the wait time at power-on or after changing the
power unit energization setting even if the activation of the output relay is disabled.
-----------------------------------------------------------------------------------------------------------------------------
Содержание KP3000GS
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 4: ......
Страница 5: ......
Страница 19: ......
Страница 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Страница 23: ......
Страница 197: ......
Страница 198: ...KP3000GS 179 5 Description of Screen and Menu 5 1 Screen Configuration 180 5 2 Menu Composition 184...
Страница 207: ......
Страница 208: ...KP3000GS 189 6 Remote Control 6 1 Communication Interface 190 6 2 Remote Local State Switching 199...
Страница 219: ......
Страница 224: ...KP3000GS 205 8 Troubleshooting 8 1 Error Messages and Error Handling 206 8 2 When a Failure is Suspected 213...
Страница 243: ......
Страница 275: ......
Страница 277: ......
Страница 278: ......
Страница 279: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...