10.19 Simulation
KP3000GS
239
10.19 Simulation
This allows you to simulate power line abnormalities, such as blackout, voltage rise, voltage drop,
abrupt phase change, and abrupt frequency change. It can be used only for AC, sine wave, and
AC+DC-INT. Note that it does not support IEC or other standards test.
KP3000GS
Number of memories
5 (non-volatile)
Number of steps
6 (Initial, Normal 1, Trans 1, Abnormal, Trans 2, Normal 2)
Setting range of Step
Time
0.0010 s to 999.9999 s (0 s is available only for the Transition Step)
Parameter Output
range
(The above item is common within the Simulation)
AC voltage
Frequency
Waveform (sine wave only)
Start Phase (excluding the Transition Step)
Stop Phase (excluding the Transition Step)
Synchronous step output (2bit)
Trigger output
Repeat count (1 to 9999 times or infinite)
初期
定常1
移行1
異常
移行2
定常2
初期
トリガ設定例
ステップ同期1出力例
ステップ同期2出力例
Initial
Initial
Normal1
Normal2
Trans1
Trans2
Abnormal
Trigger setting example
Step sync 1 output example
Step sync 2 output example
Simulation control
Start
Stop
Содержание KP3000GS
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 4: ......
Страница 5: ......
Страница 19: ......
Страница 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Страница 23: ......
Страница 197: ......
Страница 198: ...KP3000GS 179 5 Description of Screen and Menu 5 1 Screen Configuration 180 5 2 Menu Composition 184...
Страница 207: ......
Страница 208: ...KP3000GS 189 6 Remote Control 6 1 Communication Interface 190 6 2 Remote Local State Switching 199...
Страница 219: ......
Страница 224: ...KP3000GS 205 8 Troubleshooting 8 1 Error Messages and Error Handling 206 8 2 When a Failure is Suspected 213...
Страница 243: ......
Страница 275: ......
Страница 277: ......
Страница 278: ......
Страница 279: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...