10.26 External Control I/O
KP3000GS
245
10.26 External Control I/O
When you enabled the external control from the menu, you can control this product by using the
external signal (or non voltage contact). The state output is always on. The detection and state
switching is done at 10 ms-cycle. If the Remote state is achieved by the external interface, a
control input is ignored.
Item
Description
Factory
default
Control
input
External control input
Enabled or Disabled
Input level
TTL
*38
Nondestructive maximum input
+10 V /
5 V
Input impedance
Pull-up to +5 V via 47 k
Control Output Off
Falling Off
Output On
Falling On
Sequence start/resume
*39 Falling Start
Stop of sequence
*39 Falling Stop
Sequence is in the Hold status
Falling Hold
Sequence branch 1
Falling Branch start
Sequence branch 2
Falling Branch start
Memory recall (+ compile)
*40 Falling Recall
Memory specification 1
Specify 0 to 3
(Equivalent to memory 1 to 4, respectively)
Memory specification 2
Clear the current peak-hold value
Falling Clear
State
output
Output level
TTL
*41
Output impedance
220
Polarity *42 Polarity
reverse
Negative
Status
Power On/Off
Low: Off, High: On
Output On/Off*
Low: On, High: Off (Negative)
Low: Off, High: On (Positive)
Protection operation*
Low: Active, High: None (Negative)
Low: None, High: Active (Positive)
Limiter operation*
Low: Active, High: None (Negative)
Low: None, High: Active (Positive)
AGC/Autocal setting state*
Low: On, High: Off (Negative)
Low: Off, High: On (Positive)
Software busy*
Low: Busy, High: Ready (Negative)
Low: Ready, High: Busy (Positive)
Output range
Low: 200 V, High: 100 V
Sequence operation
Step sync 1
*39
High level or Low level
Sequence operation
Step sync 2
*39
Trigger Positive
or negative
Termina
l
D-sub 25-pin multi-connector (rear panel, female, M2.6 screw)
*38:
Low
: 0.8 V or lower, High: 2.6 V or higher, the chassis potential.
*39: Sequence start and stop of the control input are effective for the Simulation as well. Also, step
sync output 1 and 2 of the control output are effective for the Simulation as well.
*40: The memory recall input of the control input recalls the setting memory for the Normal
(Continuous), the Sequence Memory for the Sequence, and the Simulation Memory for the
Simulation. For the Sequence and Simulation, the compile data is also included.
*41: Low: 0.4 V or lower, High: 2.7 V or higher, the chassis potential.
*42: Polarity reverse (collective) is available only for the state with *.
Содержание KP3000GS
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 4: ......
Страница 5: ......
Страница 19: ......
Страница 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Страница 23: ......
Страница 197: ......
Страница 198: ...KP3000GS 179 5 Description of Screen and Menu 5 1 Screen Configuration 180 5 2 Menu Composition 184...
Страница 207: ......
Страница 208: ...KP3000GS 189 6 Remote Control 6 1 Communication Interface 190 6 2 Remote Local State Switching 199...
Страница 219: ......
Страница 224: ...KP3000GS 205 8 Troubleshooting 8 1 Error Messages and Error Handling 206 8 2 When a Failure is Suspected 213...
Страница 243: ......
Страница 275: ......
Страница 277: ......
Страница 278: ......
Страница 279: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...