10. Specifications
PROGRAMMABLE AC/DC POWER SOURCE
236
10.15 Remote Sensing
This switches the voltage used for measurement. When the remote sensing is on, the sensing input
terminal voltage is used. When it is off, the output terminal voltage is used.
By combining with AGC or Autocal, a voltage drop due to wiring to the load can be compensated.
When the remote sensing is on, the output voltage detection point corrected by the AGC or
Autocal function is switched to the sensing input terminal. When AGC or Autocal is off, only the
detection voltage used for measurement display is switched.
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected.
Measurement
voltage, power, power factor
AGC/Autocal
Off
On
On
Use the sensing input terminal voltage
Not active
Active
Off
Use the output terminal voltage
Not active
(factory default)
Active
10.16 AGC
When the AGC (Automatic Gain Control) is on, the detection point voltage is always measured,
and the output voltage is continuously corrected so that its effective value is equal to the output
voltage setting value. The fluctuation of the detection point voltage can be suppressed even when
the load is fluctuated. The detection point can be switched between the sensing input terminal
(remote sensing on) and the output terminal (remote sensing off).
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected. It cannot be
selected when the Autocal is set to on.
KP3000GS
Response time
Within 100 ms (typ.)
(DC/50 Hz/60 Hz, at the rated output voltage)
Operation range The output voltage setting is 8 V or higher
Calibration
range
Within
10 % (difference between the output voltage and measured value)
The output voltage should be within the allowed voltage setting range of the
product.
Accuracy
Within
0.5 V /
1.0 V
(in the case of DC or 40 Hz to 550 Hz, 50 V or higher output voltage, resistance
load, the output current is the maximum current or less)
Содержание KP3000GS
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 4: ......
Страница 5: ......
Страница 19: ......
Страница 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Страница 23: ......
Страница 197: ......
Страница 198: ...KP3000GS 179 5 Description of Screen and Menu 5 1 Screen Configuration 180 5 2 Menu Composition 184...
Страница 207: ......
Страница 208: ...KP3000GS 189 6 Remote Control 6 1 Communication Interface 190 6 2 Remote Local State Switching 199...
Страница 219: ......
Страница 224: ...KP3000GS 205 8 Troubleshooting 8 1 Error Messages and Error Handling 206 8 2 When a Failure is Suspected 213...
Страница 243: ......
Страница 275: ......
Страница 277: ......
Страница 278: ......
Страница 279: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...