4.13 Using Autocal Function
KP3000GS
147
3.
Use the [Rmt Sense] item to set the Remote sensing function to on or off, if necessary.
4.
Close the window.
4.13 Using
Autocal
Function
The Autocal (Automatic Calibration) function calculates the ratio (calibration factor) of the output
voltage measured value (effective value) versus the output voltage setting value when this function
is turned on, and multiplies the gain of the output amplifier by the calculated value, intending to
match the output voltage with the setting value. The calculated calibration factor is used as a fixed
value until the Autocal function is turned off. Therefore, even when the Autocal function is on, the
output voltage fluctuation may occur if the load is fluctuated. When the Autocal function is set to on,
the icon
is displayed.
As the detected part of the output voltage calibrated by the Autocal function, either the sensing input
terminal (the Remote sensing function is on) or the output terminal (the Remote sensing function is
off) can be selected. Combining the Remote sensing function with the Autocal function, you can
compensate the voltage drop due to wiring to the load.
The Autocal function can be set to on in the AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA
modes. However, for the AC mode, it can be set to on only when the waveform is a sine wave.
When the Autocal function is on, the waveform cannot be changed to other than a sine wave. When
the mode is changed to ACDC mode, when the signal source is changed to EXT or ADD, or when
the Sequence or Simulation function is selected, the Autocal function is forcibly set to off. When the
AGC function is on, the Autocal function cannot be used.
For the specification of the Autocal function, see
10.17
.
Содержание KP3000GS
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 4: ......
Страница 5: ......
Страница 19: ......
Страница 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Страница 23: ......
Страница 197: ......
Страница 198: ...KP3000GS 179 5 Description of Screen and Menu 5 1 Screen Configuration 180 5 2 Menu Composition 184...
Страница 207: ......
Страница 208: ...KP3000GS 189 6 Remote Control 6 1 Communication Interface 190 6 2 Remote Local State Switching 199...
Страница 219: ......
Страница 224: ...KP3000GS 205 8 Troubleshooting 8 1 Error Messages and Error Handling 206 8 2 When a Failure is Suspected 213...
Страница 243: ......
Страница 275: ......
Страница 277: ......
Страница 278: ......
Страница 279: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...